Remote Control Commands
R&S
®
SMB100A
249Operating Manual 1407.0806.32 ─ 13
For an overview of the settings affected by the preset function, see chapter 5.2.2, "Default
Instrument Settings - Preset Key", on page 80.
Example:
SYST:PRES
All instrument settings (also those that are not currently active) are
reset to their default values.
Usage: Setting only
:SYSTem:FPReset
The command triggers an instrument reset to the original state of delivery.
Note: Since Factory Preset resets the Remote Channel and network settings to the
default values, executing Factory Preset via remote control terminates the connection to
the instrument, if these settings had been configured to values different to the default
ones!
The factory preset function resets nearly all instrument settings. In addition to the regular
preset by means of the PRESET key, a "Factory Preset" resets also the following values:
Reference frequency settings ("Ref Oscillator" menu)
Power on settings ("Level/EMF" menu)
Network settings including hostname ("Setup" menu)
Remote Channel settings including GPIB address ("Setup" menu)
Start/Stop Gui Update ("Setup" menu)
Display and keyboard settings ("Setup" menu).
To maintain security, password settings and all settings protected by these passwords
like disabled USB and LAN connections are not changed.
Not affected by the "Factory Preset" are also user data, lists or instrument settings files,
created for example by means of the Save/Recall function.
Example:
SYST:FPR
all instrument settings (also those that are not currently active) are
reset to the factory values.
Usage: Event
Manual operation: See "Factory Preset" on page 102
7.3 CALibration Subsystem
The CALibration system contains the commands for adjustment. Adjustment is triggered
by the query commands. The response "0" indicates error-free adjustment, and the
response "1" means that an error occurred during adjustment.
:CALibration<hw>:ALL[:MEASure]?.................................................................................250
:CALibration<hw>:FMOFfset[:MEASure]?.........................................................................250
:CALibration<hw>:FREQuency[:MEASure]?......................................................................250
:CALibration<hw>:LEVel[:MEASure]?...............................................................................251
CALibration Subsystem