Remote Control Commands
R&S
®
SMB100A
251Operating Manual 1407.0806.32 ─ 13
:CALibration<hw>:LEVel[:MEASure]? [, <Force>]
The command starts all adjustments which affect the level.
The acquired correction values improve the settling time and the signal quality.
Query parameters:
<Force> string
*RST: force
Return values:
<Measure> 0 | 1
Example:
CAL:LEV:MEAS?
starts adjustments for maximum level accuracy.
Response: "0"
adjustment has been performed successfully.
Usage: Query only
Manual operation: See "Adjust Level" on page 82
:CALibration<hw>:LEVel:EXTern:DATA <Data>
Selects the data used for of external level correction.
By default the instrument uses correction data obtained in the factory before delivery. In
addition, customer data can be used for external level correction. The customer data is
obtained using a R&S NRP power sensor. External level correction is a protected function
(see service manual, chapter 2, "Adjustment").
Parameters:
<Data> FACTory | CUSTomer
*RST: FACTory
Example:
CAL:LEV:EXT:DATA FACT
selects the use of the data aquired at the factory for external level
correction.
Manual operation: See "Adjustment Data" on page 82
:CALibration<hw>:ROSCillator[:DATA] <Data>
Sets the calibration value for the custom defined external adjustment.
Parameters:
<Data> integer
Range: 0 to INT_MAX
*RST: 0
CALibration Subsystem