Remote Control Commands
R&S
®
SMB100A
250Operating Manual 1407.0806.32 ─ 13
:CALibration<hw>:LEVel:EXTern:DATA............................................................................251
:CALibration<hw>:ROSCillator[:DATA].............................................................................251
[:SOURce]:CALibration:STEReo:ANALog[:MEAS]?...........................................................252
:CALibration<hw>:ALL[:MEASure]? [, <Force>]
Starts all internal adjustments that do not require external measurement equipment.
Query parameters:
<Force> string
Return values:
<Measure> select
Example:
CAL:ALL:MEAS?
starts the adjustment of all functions for the entire instrument.
Response: "0"
adjustment has been performed successfully.
Usage: Query only
Manual operation: See "Adjust All" on page 82
:CALibration<hw>:FMOFfset[:MEASure]?
The command starts all adjustment for the FM/PhiM modulator.
Return values:
<Measure> 0 | 1
Example:
CAL:FMOF?
starts the adjustments for the FM/Phim modulator.
Response: "0"
the adjustments have been performed successfully
Usage: Query only
Manual operation: See "Adjust FM Offset" on page 175
:CALibration<hw>:FREQuency[:MEASure]?
The command starts all adjustments which affect the frequency.
Return values:
<Measure> 0 | 1
Example:
CAL:FREQ:MEAS?
starts the adjustments for maximum frequency accuracy.
Response: "0"
the adjustments have been performed successfully.
Usage: Query only
Manual operation: See "Adjust Synthesis" on page 82
CALibration Subsystem