Return values:
<Points> integer
Range: 0 to INT_MAX
*RST: 0
Example:
MMEM:CDIR '/var/user/lists/pulsetrain'
selects the directory for the pulse train files.
PULM:TRA:SEL ‘P_INCR'
selects P_INCR for editing. P_INCR is created if it does not yet
exist.
PULM:TRA:ONT:POIN?
queries the number of frequency values in P_INCR
Response: 7
P_INCR has 7 ontime entries.
Usage: Query only
Options: R&S SMB-K27 (Pulse Train)
[:SOURce<hw>]:PULM:TRAin:REPetition <Repetition>
Sets the number of repetitions for each ontime/offtime value pair.
*RST does not affect data lists.
Tip: "0" ignores the corresponding value pair in the pulse train. Thus, you can individually
omit value pairs without deleting them from the table.
Parameters:
<Repetition> Repetition#1{, Repetition#2, ...}
Range: 0...65535
Example:
MMEM:CDIR 'var/lists/pulsetrain'
selects the directory for the pulse train files.
PULM:TRA:SEL 'P_INCR'
selects P_INCR for editing. P_INCR is created if it does not yet
exist.
PULM:TRA:ONT 10ns,15ns,40ns,...
specifies the ontime values in P_INCR. If the list already contains
data, it is overwritten.
PULM:TRA:OFFT 10ns,15ns,40ns,...
specifies the offtime values in P_INCR. If the list already contains
data, it is overwritten.
PULM:TRA:REP 10,15,40,...
specifies the number of repetitions for each value pair.
Options:
R&S SMB-K27 (Pulse Train)
Manual operation: See "Edit Pulse Train Data" on page 198