Remote Control Commands
R&S
®
SMB100A
397Operating Manual 1407.0806.32 ─ 13
:SYSTem:COMMunicate:HISLip:RESource?
Queries the VISA resource string . This string is used for remote control of the instrument
with HiSLIP protocol.
Return values:
<Resource> string
Example:
SYST:COMM:HISL:RES
queries the VISA resource string.
Response: TCPIP::192.1.2.3::hislip0::INSTR
Usage: Query only
Manual operation: See "Visa Resource Strings" on page 96
:SYSTem:COMMunicate:NETWork[:COMMon]:DOMain <Domain>
Sets the domain of the network.
Parameters:
<Domain> string
Example:
SYST:COMM:NETW:COMM:DOM ABC.DE
sets the domain of the network.
:SYSTem:COMMunicate:NETWork[:COMMon]:HOSTname <Hostname>
Sets the individual computer name of the R&S Signal Generator. The entry is activated
only after sending command SYST:COMM:NETW:COMM:SET.
Note: The hostname can only be changed after deactivating protection level 1 (command
SYSTem:PROTect<n>:STATe OFF, password)
Parameters:
<Hostname> string
Example:
SYST:COMM:NETW:HOST ‘SIGGEN’
sets the individual computer name of the R&S Signal Generator.
Example:
SYST:COMM:NETW:COMM:SET ON
activates the change of the hostname.
Manual operation: See "Hostname" on page 91
:SYSTem:COMMunicate:NETWork[:COMMon]:WORKgroup <Workgroup>
Sets the individual workgroup name of the instrument.
Parameters:
<Workgroup> string
Example:
SYST:COMM:NETW:COMM:WORK 'TEST_09'
sets the workgroup name 'TEST_09'.