Remote Control Commands
R&S
®
SMB100A
348Operating Manual 1407.0806.32 ─ 13
The specified value range is therefore only effective if :SOURce:POWer:OFFSet is set
to 0. The value range for other OFFset values can be calculated using the following
formula:
Minimum level + OFFSet ... Maximum level + OFFSet
Parameters:
<Start> float
Minimum level ... Maximum level
The value range for the level setting varies according to the instru-
ment model. The values are given in the data sheet.
Increment: 0.01 dB
*RST: -30 dBm
Default unit: dBm
Example:
POW:STAR -20 dBm
sets the start level for the level sweep to -15 dBm for RF output A.
Manual operation: See "Start Level - Level Sweep" on page 158
[:SOURce<hw>]:POWer:STEP[:INCRement] <Increment>
The command sets the step width for the level setting if UP and DOWN are used as the
level values and variation mode :SOUR:POW:STEP:MODE USER is selected. The com-
mand is linked to setting "Variation Step" for manual control, i.e. the command also sets
the step width of the rotary knob for "Variation Active On".
Parameters:
<Increment> float
Range: 0 dB to 100 dB
Increment: 0.01 dB
*RST: 1 dB
Default unit: dB
Example:
SOURce:POWer:STEP:INCRement 2
or
POW:STEP 2
sets the step width for entering the RF level to 2 dB.
Manual operation: See "Variation Step" on page 127
[:SOURce<hw>]:POWer:STEP:MODE <Mode>
This command activates (USER) or deactivates (DECimal) the user-defined step width
used when varying the level value with the level values UP/DOWN. The command is linked
to setting "Variation Active" for manual control, i.e. the command also activates/deacti-
vates the user-defined step width used when varying the level value with the rotary knob.
Parameters:
<Mode> DECimal | USER
*RST: DECimal
SOURce Subsystem