Instrument Function
R&S
®
SMB100A
158Operating Manual 1407.0806.32 ─ 13
Example:
SOUR:SWE:POW:MODE AUTO
TRIG:PSW:SOUR SING
SOUR:POW:MODE SWE
SOUR:SWE:EXEC
Remote command:
[:SOURce<hw>]:SWEep:POWer:EXECute on page 385
:TRIGger<hw>:PSWeep[:IMMediate] on page 409
:TRIGger<hw>[:SWEep][:IMMediate] on page 411
Start Level - Level Sweep
Sets the start level.
Remote command:
[:SOURce<hw>]:POWer:STARt on page 347
Stop Level - Level Sweep
Sets the stop level.
Remote command:
[:SOURce<hw>]:POWer:STOP on page 349
Current Level - Level Sweep
Displays the current level.
If "Step" is set, the level for the next level step of the sweep is entered here.
Remote command:
[:SOURce<hw>]:POWer:MANual on page 344
Shape - RF Level Sweep
Selects the cycle mode for a sweep sequence (shape).
"Sawtooth"
One sweep runs from the start level to the stop level. The subsequent
sweep starts at the start level again, i.e. the shape of sweep sequence
resembles a sawtooth.
"Triangle"
One sweep runs from start to stop level and back, i.e. the shape of the
sweep resembles a triangle. Each subsequent sweep starts at the start
level again.
Remote command:
[:SOURce<hw>]:SWEep:POWer:SHAPe on page 386
Step - Level Sweep
Sets the step width for the individual sweep steps. This entry is effective for all sweep
modes.
With the level sweep, the logarithmic step width is a constant fraction of the current level.
This fraction is added to the current level. The logarithmic step width is entered in dB.
Remote command:
[:SOURce<hw>]:SWEep:POWer:STEP[:LOGarithmic] on page 387
RF Block