Remote Control Commands
R&S
®
SMB100A
346Operating Manual 1407.0806.32 ─ 13
[:SOURce<hw>]:POWer:SPC:CRANge <PowCntrlCRange>
Defines the capture range of the power control system.
Within the range:
Target Level +/- Catch Range
the power control locks and tries to achieve the target level. Readings outside the range
are not considered.
Parameters:
<PowCntrlCRange> float
Range: 0 to 50
Increment: 0.01
*RST: 30
Default unit: dB
Example:
POW:SPC:CRAN 15
sets the capture range to +/- 15 dB.
Manual operation: See "Catch Range +/-" on page 133
[:SOURce<hw>]:POWer:SPC:DELay <PowCntrlDelay>
Defines a waiting period between the level adjustment of the generator and the next
measurement of the power sensor.
Parameters:
<PowCntrlDelay> integer
Range: 0 to 1000
Increment: 1
*RST: 0
Example:
:POW:SPC:DEL 2 ms
the sensor starts the next reading 2 ms after the level adjustment.
Manual operation: See "Delay Time" on page 134
[:SOURce<hw>]:POWer:SPC:PEAK <PowCntrlPeak>
Activates power control by means of the peak power values, provided the power sensor
supports this function.
Parameters:
<PowCntrlPeak> 0 | 1 | OFF | ON
*RST: 0
Example:
POW:SPC:PEAK ON
uses the measured peak power for power control.
Manual operation: See "Use Peak Power" on page 134
SOURce Subsystem