Remote Control Commands
R&S
®
SMB100A
345Operating Manual 1407.0806.32 ─ 13
[:SOURce<hw>]:POWer:MODE <Mode>
The command sets the instrument operating mode and therefore also the commands
used to set the output level.
Parameters:
<Mode> CW | FIXed | SWEep
CW|FIXed
The instrument operates at a constant level. CW and FIXed are
synonyms. The output level is set with :SOURce:POWer.
SWEep
The instrument operates in SWEep mode.
The level is set using the commands SOURce:POWer:STARt;
STOP; MANual.
LIST
The instrument processes a list of frequency and level settings.
The List mode settings are made in the SOURce:LIST Subsys-
tem subsystem.
The setting SOURce:POWer:MODE LIST also sets the command
SOURce:FREQuency:MODE automatically to LIST.
*RST:
CW
Example:
POW:MODE SWEep
sets the SWEep mode. The settings under SOURce:POW:STARt;
STOP; MANual become effective.
Manual operation: See "State - Level Sweep" on page 155
[:SOURce<hw>]:POWer:POWer <Power>
Sets the RF level of the RF output connector.
The level entered with this command corresponds to the level at the RF output, i.e. any
offset entry is not taken into consideration.
Note: The SCPI command [:SOURce<hw>]:POWer[:LEVel][:IMMediate][:
AMPLitude] sets the level of the "Level" display, i.e. the level containing offset.
Parameters:
<Power> Minimum level ... Maximum level
The value range for the level setting varies according to the instru-
ment model.
The values are given in the data sheet.
Increment: 0.01 dB
*RST: -30 dBm
Default unit: dBm
Example:
SOUR:POW:POW 15
sets the RF level at output to 15 dBm.
Manual operation: See "Amplitude" on page 124
SOURce Subsystem