Remote Control Commands
R&S
®
SMB100A
309Operating Manual 1407.0806.32 ─ 13
Example:
FREQ:STOP 2 GHz
sets the stop frequency for the frequency sweep to 2 GHz.
FREQ:STAR 1 MHz
sets the start frequency for the frequency sweep to 1 MHz.
Manual operation: See "Stop Freq - Frequency Sweep" on page 152
[:SOURce<hw>]:FREQuency:STEP[:INCRement] <Increment>
The command sets the step width for the frequency setting if the frequency values
UP/DOWN are used and variation mode SOUR:FREQ:STEP:MODE USER is selected. The
command is linked to "Variation Step" for manual control, i.e. the command also sets the
step width of the rotary knob for "Variation Active" on.
Parameters:
<Increment> float
Range: 0 Hz to RFmax - xx kHz (depending on model)
Increment: 0.01 Hz
*RST: 1 MHz
Example:
FREQ:STEP 50 kHz
sets the step width for the frequency setting to 50 kHz.
Manual operation: See "Variation Step - RF Signal" on page 119
[:SOURce<hw>]:FREQuency:STEP:MODE <Mode>
This command activates (USER) or deactivates (DECimal) the user-defined step width
used when varying the frequency value with the frequency values UP/DOWN. The com-
mand is linked to the command "Variation Active" for manual control, i.e. the command
also activates/deactivates the user-defined step width used when varying the frequency
value with the rotary knob.
Parameters:
<Mode> DECimal | USER
*RST: DECimal
Example:
FREQ:STEP 50 kHz
sets the step width for the frequency setting to 50 kHz.
FREQ:STEP:MODE USER
actives this step width for frequency variation with the rotary knob
(manual control) and with frequency values UP/DOWN (remote
control).
Manual operation: See "Variation Active - RF Signal" on page 118
SOURce Subsystem