Instrument Function
R&S
®
SMB100A
135Operating Manual 1407.0806.32 ─ 13
If user correction is activated, the "UCOR" display (User Correction) is shown in the
header together with the "Level" display. The RF output level is the sum of both values.
"Level" + "UCOR" = Output level
If activated, user correction is effective in all operating modes.
User Correction Menu
To open the "User Correction" menu, select "RF > Configure > User Correction" or use
the MENU key under "RF".
The combined menu "ALC/UCOR" is divided into the several sections.
User Correction Settings
The "User Correction" settings are set in the most lower section of the combined dialog;
this section is used to activate/deactivate user correction, and to create, select and acti-
vate the lists.
The upper section provides access to the automatic level control settings, see chap-
ter 5.3.5.4, "Automatic Level Control - ALC", on page 128.
State - User Correction
Activates/deactivates user correction.
The "UCOR" status message appears in the frequency and level display.
Remote command:
[:SOURce<hw>]:CORRection[:STATe] on page 299
User Correction Value - User Correction
Indicates the current value for level correction.
Remote command:
[:SOURce<hw>]:CORRection:VALue? on page 299
User Cor. Data - User Correction
Calls the "File Select" menu for selecting and creating a list or the "File Manager".
Remote command:
[:SOURce]:CORRection:CSET:DELete on page 294
[:SOURce<hw>]:CORRection:CSET[:SELect] on page 298
RF Block