Remote Control Commands
R&S
®
SMB100A
253Operating Manual 1407.0806.32 ─ 13
Example:
DIAG:BGIN
Queries the instrument configuration.
returns the data of all available modules.
DIAG:BGIN? 'MBRD'
Queries the configuration of the motherboard.
Response: MBRD 1141.3501.02 1.5.3 100023
Module motherboard with stock number 1141.3501.01 has revi-
sion 1.5.3 and serial number 100023.
Usage: Query only
Manual operation: See "Assembly" on page 84
:DIAGnostic<hw>:BGINfo:CATalog?
The command queries the names of the assemblies available in the instrument.
A complete list of all assemblies is returned (the various entries are separated by com-
mas). The length of the list is variable and depends on the instrument equipment con-
figuration.
Return values:
<Catalog> string
Example:
DIAG:BGIN:CAT
queries the names of the assemblies.
Usage: Query only
:DIAGnostic:INFO:OTIMe?
The command queries the number of operation hours.
Return values:
<OTIMe> float
Example:
DIAG:INFO:OTIM
queries the operation hours.
Response: 100023
The instrument was operated for 100023 hours up to now.
Usage: Query only
Manual operation: See "Operation Time / h" on page 83
:DIAGnostic:INFO:POCount?
The command queries the number of power-on events.
Return values:
<Pocount> float
DIAGnostic Subsystem