Remote Control Commands
R&S
®
SMB100A
367Operating Manual 1407.0806.32 ─ 13
Parameters:
<Value> integer
Range: 0 to 4095
Increment: 1
*RST: 1400
Example:
ROSC:ADJ:VAL 1400
sets the adjustment value to 1400.
Manual operation: See "Adjustment DAC Value" on page 121
[:
SOURce]:ROSCillator[:INTernal]:ADJust[:STATe] <State>
The command determines whether the calibrated (OFF) or a user-defined (ON) adjust-
ment value is used for fine adjustment of the frequency. With STATe ON, the instrument
is no longer in the calibrated state. However, the calibration value is not changed and the
instrument resumes the calibrated state after switching the Adjustment State to Off.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: --
Example:
ROSC:SOUR INT
selects the internal source.
ROSC:ADJ ON
activates use of a user-defined adjustment value.
ROSC:ADJ:VAL 1400
sets the adjustment value to 1400.
Manual operation: See "Adjustment Active - Reference Oscillator" on page 121
[:SOURce]:ROSCillator:SOURce <Source>
Selects the reference frequency source.
Parameters:
<Source> INTernal | EXTernal | ELOop
INTernal
The internal reference oscillator is used.
EXTernal
An external reference signal is used. It must be input at the REF
IN connector at the rear of the instrument.
The instrument is informed of the frequency of the external refer-
ence signal by means of the command [:SOURce]:
ROSCillator:EXTernal:FREQuency.
*RST: ---
SOURce Subsystem