Operation
A5-6 ENVI-MAG Manual
Press the “SETUP” and “ON” keys simultaneously and select configuration
2-Drum Locations. This is described fully in “Configuration menu” on
page A4-3
Check your setup
1. Set up Line and Station Number and Separations as shown in the example
under “Line and Station setup” on page A4-7.
2. Do the setup as shown under “Sample total-field setup” on page A4-9.
Start survey
3. Proceed to your first location.
4. Press the “START” key.
5. Proceed immediately with your desired pace toward the West.
6. Press the “NEXT” key when the sensor passes the first station peg. If you
trigger the station marker at the time that you yourself pass the station,
rather than the sensor, you will likely introduce a herring-bone pattern into
the survey data. Observe how the Station Number decrements.
7. Repeat the previous step at each peg.
8. At Station 50 W, Press “STOP” right after you have pressed “NEXT”.
9. You can now inspect your data graphically (before you press “ESCAPE”) by
pressing the” NUM/GRAPH” key. This will show you up to the last 178
readings on the line.
10. Press the “ESCAPE” key.
11. Move the cursor to ST: SEP:
12. Enter! 10.
13. Move the cursor to LN:
14. Press the + key. The Line number increments to 18.
15. Proceed to Line 18 N, Station 50 W.
16. Press the “START” key.
17. Proceed immediately walking to the East.