EasyManuals Logo

SCINTREX ENVI-MAG User Manual

SCINTREX ENVI-MAG
265 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #118 background imageLoading...
Page #118 background image
Operation
A5-10 ENVI-MAG Manual
18. At the Main operating display move to cursor to ST: SEP:
19. Enter! 10.
20. Move the cursor to LN:
21. Press the “+” key to increment the line number to 18 N.
22. Take a reading and proceed along line 18 N toward the East.
23. At 10 W again, before you press the “RECORD” key, Press the “NOTE” key,
and then the “1” key.
24. After you have finished the line, move to Line 20 N and proceed to 50 E.
25. At the Main operating display move the cursor to ST: SEP:
26. Enter! -10.
27. Move the cursor to LN:
28. Press the “+” key to increment the line number to 20 N.
29. Proceed with Line 20 N.
Automating your measurements
There are two items which can be automated after each measurement:
data recording,
altering of the Station Number can take place automatically.
Caution: Please note that repeat measurements will alter the Station Number and
possibly disrupt the correlation between location and data.
To enable both features do the following:
1. In the Instrument setup, move the cursor to AUTO ST.INC.:
2. Select! YES
3. In the mag setup, move the cursor to AUTO RECORD:
4. Select! YES
To operate as a total-field magnetometer in the Stop-and-Go mode, make the
following change in the
mag setup: screen, MODE: T-FLD