Setup
A4-14 ENVI-MAG Manual
Sample total-field setup
(WALKMAG or WALKGRAD)
The following steps prepare the ENVI-MAG for:
▲ a total-field survey in the WALKMAG or WALKGRAD modes,
▲ with 0.5 second reading period,
▲ automatic tuning,
▲ manual station increment.
From the Main operating display proceed as follows:
1. Press the “SETUP” key and the INSTRUMENT SETUP display appears:
MAG SETUP: ->menu AUTO ST.INC.: yn
LINK REC/START: yn
CYCLE REPEAT: yn
CYCLE DELAY: yn
ERASE MEMORY: yn
HEATER: yn
Chg?:ENT. BATT:bbb
2. Carry out steps 2 through 9 (as shown on page A4-13) for setting up the
instrument, if you have not already done so.
3. Move the cursor to MAG SETUP.
4. Press “ENTER” and the MAG SETUP display appears.
MODE: mmm CYCLE TIME: ttt
DURATION: ddd CHART SCALE: sss
TUNE FIELD: fffff AUTO TUNE: yn
BASE CORRECT: yn BASE FIELD: bbbbb
TIE CORRECT: yn TIE-MODE: lll
REMOTE: yn AUTO RECORD: yn
Chg?:+- BAT:bbb
5
instrument setup
3
mag setup