Operating displays
ENVI-MAG Manual A3-23
Instrument
Instrument setup display
This display enables you to configure the basic data acquisition portion of
the ENVI-MAG. This display is accessible with the “SETUP” key.
instrument setup
MAG SETUP: ->menu AUTO ST.INC.: yn
LINK REC/START: yn
CYCLE REPEAT: yn
CYCLE DELAY: ddd
ERASE MEMORY: yn
HEATER: yn
Chg?:ENT. BATT:bbb
Figure A-23 ENVI-MAG instrument setup display.
SEP
The line or station separation.
The range is from 0 to 99999 with a decimal point, as
required, and can either be positive (+) or negative (-). It may
not necessarily represent the reading separation.
See the “NEXT” key and AUTO. St. INC. for more information.
When the separation is positive, the station number gets
incremented by the separation.
When the separation is negative, the station number gets
decremented by the separation.
Note: a
negative
(W,S,-) station, when incremented,
becomes
less
negative, and vice versa.
LN
The line number, along which the measurements take place.
It consists of the numeric part (
llll.l) and the
directional part [
d].
The range is from 0 to 99999 with a decimal point as
required.
The direction allows you to enter the geographical
direction or the Cartesian co-ordinates as one of:
N,E,S,W, + or -.
Parameter Description
5
instrument setup