Instrument
A3-24 ENVI-MAG Manual
Parameter Description Default
MAG
SETUP
Displays the magnetometer specific setup menu.
Pressing the “ENTER” key gets the next menu.
AUTO
ST.INC.
A “yes/no” toggle that controls the automatic
station increment (decrement).
no for the base station, walking or manual
type of operation.
yes for semi-automatic operation.
no
LINK
REC/
START
Link Record/Start
This setting should be left at
no for
magnetometer only operation.
no
CYCLE
REPEAT
Determines whether the magnetometer will
take one reading only, or will continuously take
readings at the time interval specified by the
DURATION parameter in the Mag. Setup.
This is
not applicable for Base Station
operation, which is set in the Mag Setup.
no
CYCLE
DELAY
The delay between readings when the cycle
repeat feature is enabled.
It affects the total-field magnetometer or
gradiometer only.
The base-station repetition rate is controlled in
the Mag. Setup under Cycle Time.
The cycle delay is the time between the end of one
reading to the beginning of the next reading. This
delay is useful when the data density is thought
to be too large, or in hilly terrain where the data
density would increase when walking uphill.
0
ERASE
MEMORY
Allows clearing of the data memory.
A request for confirmation will be displayed
before the actual erasure takes place.
NOTE: The setup parameters remain intact.
no
HEATER
Enables or disables the LCD display heater.
If the temperature is above -15 degree C,
enabling this parameter has no effect.
no