EasyManua.ls Logo

SCINTREX ENVI-MAG - Search Mode Data Collection Displays

SCINTREX ENVI-MAG
265 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating displays
ENVI-MAG Manual A3-21
Instrument
Search mode data collection displays
The display screen during data collection is as follows:
Figure A-20
Numeric data collection display during Search mode
You can push the “NUM/GRA” key to display a profile of up to the last 178
data points collected at any time in the Search mode. The display appears as
follows:
-*MAG*-
TOTAL
mmmmm.m
LN:
xxx.x E
ST:
xxx.x N
Figure A-21 Graphical display during Search mode.
If you stop the survey, you can press the “NEXT” key to pop-up the graphical
display of the precession decay signal of the last reading. Very sharp, rapid
decays indicate possible poor signal conditions. Please see Figure A-18 on
page A3-18 for an example of the signal quality display.
Note: The values that you see for the line and station in the numeric display are
the last ones entered in a previous mode. The search mode does not use
these for any purpose, so please ignore them.
00 -* MAG *- 13:45:30
TOTAL NOISE GRADIENT LN:xxx.x E
59875.2 1.5 110.2 ST:xxx.x N
59623.7 0.9 62.9 xxx.x N
59590.3 0.8 15.1 xxx.x N
59581.4 1.9 2.3 xxx.x N
MEMF: 88%
BATT: 134
Reading
Previous readings
Free memory
Battery strength
Magnetic value
Noise
4
wt

Table of Contents