EasyManua.ls Logo

SCINTREX ENVI-MAG - Upper Sensor

SCINTREX ENVI-MAG
265 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Assembly
ENVI-MAG Manual A2-7
Preparation
Upper sensor
12. Select the upper sensor–it only has a pair terminals at the 12 o’clock
position. (See “Figure A-5Gradiometer sensor cabling” on page A2-7.)
13. Attach the upper sensor to the lower one by placing the shaft of the
lower sensor into the socket of the upper one while rotating slightly
counter-clockwise. When they are fully engaged, firmly twist them
clockwise against each other.
Warning: Both sensors must have their directional marks aligned in the same
manner, i.e. the N on the top sensor must be in-line with the N on the
bottom sensor.
14. Remove the cable hold-down plate of the upper sensor by unscrewing
the screws at the 2 o’clock and 4 o’clock positions.
15. Connect the white wire of the two-conductor cable coming from the
lower sensor to the terminal at the 1 o’clock position.
16. Connect the bare wire of the two-conductor cable coming from the
lower sensor to the terminal at the 11 o’clock position.
17. Re-attach the cable hold-down plate.
Figure A-5 Gradiometer sensor cabling

Table of Contents