Setup
A4-20 ENVI-MAG Manual
Sample base-station setup
The following steps prepare the ENVI-MAG as a:
magnetic base-station,
with a 2 second reading period,
and fixed tuning.
From the Main operating display proceed as follows:
1. Press the “SETUP” key and the INSTRUMENT SETUP display appears.
MAG SETUP: ->menu AUTO ST.INC.: yn
LINK REC/START: yn
CYCLE REPEAT: yn
CYCLE DELAY: yn
ERASE MEMORY: yn
HEATER: yn
Chg?:ENT. BATT:bbb
2. Follow the instructions for the Instrument setup, as shown on page A4-12,
but substitute the following.
3. Move the cursor to CYCLE REPEAT:
4. Select! no
5. Move the cursor to CYCLE DELAY:
6. Enter! 0
7. Move the cursor to MAG SETUP:
8. Press “ENTER”.