Advanced Mode
ENVI-MAG Manual A4-13
Setup
2. Move the cursor to AUTO ST.INC: (automatic station increment)
3. Select! no (Since the ENVI-MAG is reading continuously, you want the
station value to change only when you reach a station— not with every
reading!)
4. Move the cursor to LINK REC/START:
5. Select! no
6. Move the cursor to CYCLE REPEAT:
7. Select! yes (You want the instrument to continuously take readings at
the rate specified by the
DURATION parameter)
8. Move the cursor to CYCLE DELAY:
9. Enter! 250 (This value is in milliseconds, and pauses the reading cycling
by 250 ms. This is useful when you are traversing uneven ground where your
speed may vary)
10. Ignore ERASE MEMORY:
11. Move the cursor to HEATER:
12. Select! no (unless you are doing a winter survey and the ambient
temperature is below minus 15° C.)
13. Press “ESCAPE” to return to the Main operating display.