Trouble shooting
ENVI-MAG Manual A7-9
Maintenance
Reading noise is high
No sensor(s) attached Connect the sensor(s).
Tuning is incorrect 1. Set manual tuning
properly
2. Use Auto-tuning.
Sensor cable is
improperly connected
Connect properly.
Sensor cable is broken Repair sensor cable.
Improper sensor
orientation
Make sure that either
the N or S mark on
the sensor points to
magnetic North.
High gradient 1. Monitor the
Numeric Page 2
display bar chart.
2. Readings may be
impossible to make
in this situation.
External interference
(e.g. power line)
Readings may be
impossible.
No fluid in the sensor
(shake and listen
for sloshing sound)
Contact SCINTREX.
Problems Possible Causes Possible Solutions