EasyManua.ls Logo

Thermo Scientific Applied Biosystems SeqStudio Genetic Analyzer - Page 178

Thermo Scientific Applied Biosystems SeqStudio Genetic Analyzer
242 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Observation Possible cause Recommended action
Dye blobs are seen in the
sequencing data
Impurities remained in the
sample after the sample
purification. The impurities
cause dye blobs to appear in
the sequencing data.
Improve the sample purification method. See
“Sample preparation guidelines“ on page 37
for guidelines.
Extra peaks are present in the
sequencing traces
There was renaturation of the
sample.
Heat-denature the samples prepared with
fresh HiDi
Formamide, then immediately
place the samples on ice.
There is low signal. With very
low signal, the peaks are barely
visible in the baseline noise.
Check the raw data, the raw data signal
intensity, and average raw signaltonoise ratio,
then:
Increase the injection time and reinject.
Or
Remake the sample. Ensure that you are
using:
Enough sequencing template
Enough primer and/or a sufficient
concentration of primer
There is a heterozygous
insertion-deletion (het indel)
that is causing multiple peaks
to appear at the same basecall
position. The sequence can
appear "clean" for some
number of bases until the het
indel is encountered.
Examine the analyzed trace. A het indel
typically has single peaks at the 5 end, then
part-way through the trace, two peaks appear
in almost every position to the end of the trace.
This pattern occurs when one copy of the gene
has an insertion or deletion relative to the
other copy of the gene.
When aligning your sequence to a reference
sequence, a series of bases may have been
inserted or deleted in an allele. These indels
can be encountered in any number of bases
after the gene-specific priming region. To
confirm that the het indel is present in both
directions of your target, check the sequencing
in the opposite direction.
Appendix A Troubleshooting
Sample and data troubleshooting
A
178
SeqStudio
Genetic Analyzer Instrument and Software User Guide

Table of Contents