Table 17 HID run modules
Run module
Resolution
range
Approximate
run time
Sizing precision Compatible size standards
HIDAnalysis 60–470 bp
[1]
39 minutes 60–470: <0.15
GeneScan
500 LIZ
Size
Standard
[2]
GeneScan
600 LIZ
Size
Standard v2.0
[1]
Resolution Range: The range of bases over which the resolution (peak spacing interval divided by the peak width at half-max in a GS600 size
standard sample sized with a third order fit) is 1.
[2]
The GeneScan
500 LIZ
Size Standard was not included in the HID validation of the instrument.
Note: The following size standards have not been validated for use with the
instrument. A default size standard denition is not provided in the software.
·
GeneScan
350 ROX
Size Standard
·
GeneScan
400HD ROX
dye Size Standard
Appendix E Run modules
Run modules, read lengths, size ranges, and run times
E
SeqStudio
Genetic Analyzer Instrument and Software User Guide
219