The Object Name column lists the items that require signatures:
If the Object Name
lists
The object is For information, see
A dye set name The most recent dye calibration
for the dye set specified in the
plate setup.
“Sign a dye calibration or an
install run“ on page 64
An install standard
name
The most recent install run for
the application specified in the
plate setup.
“Sign a dye calibration or an
install run“ on page 64
A plate name Plate setup “Sign a plate setup“ on
page 63
The example below indicates that E-signature is
congured to require 4 signatures
when you start a run, 1 signature is complete and 3 signatures are pending:
• 2 administrator signatures for dye calibration (dye set name is listed two times):
– 1 signature that species the "Reviewed and approved" E-signature meaning
(complete, because 1 signature is required [Signatures Required=1] and 1
signature has been logged [Current=1]).
– 1 signature that species the "Checked consumables" meaning (incomplete).
• 1 administrator signature for the plate setup that species the "Reviewed and
approved" E-signature meaning (incomplete).
• 1 administrator signature for install run that species the "Reviewed and
approved" E-signature meaning (incomplete).
Chapter 4 Use the instrument with the Security, Auditing, and E‑signature (SAE) v2.0 module
E-signature requirements when you start a run
4
62