TH2829X Series Operation Manual Chapter 8 Performance and Test
When the test speed is FAST, select the value in the bottom, such as 0.1.
When in the boundary line of the frame, select a smaller value.
The basic measurement accuracy factor A is applicable to test level range of
500mVrms--1.0Vrms. When exceeding this level range, select the value of A
L
according to Table A, the current basic measurement accuracy is obtained by
calculating A
L
A.
Where, V
s
is test signal voltage.
Table A Level correction factor A
L
Test signal voltage Vs (Vrms)
Figure B shows the level correction factor A
L
in graphic method according to the
data listed in Table A.
Figure B Level correction factor A
L
curve
Table B Impedance rate factors: K
a
, K
b
msm fVZ
100
)
200
1)(
101
(
3
ms
m
fV
Z
100
)
70
1)(103.0(
9