TH2829X Series Operation Manual Chapter 10 HANDLER
159
TSDisp Set the display page as the transformer scan test page.
For example: WrtCmd(DISP:PAGE MEAS), Set the display page as the LCR
measurement display page.
Query syntax: DISPlay:PAGE?
Return format: <page name><NL^END>
<page name> can be set as the following items:
<LCR MEAS DISP> The current page is the LCR measurement display page.
<BIN No. DISP> The current page is the bin number display page.
<BIN COUNT DISP> The current page is the bin count display page.
<LIST SWEEP DISP> The current page is the list sweep display page.
<MEAS SETUP> The current page is the measurement setup page.
<CORRECTION> The current page is the correction page.
<LIMIT TABLE SETUP> The current page is the limit table setup page.
<LIST SWEEP SETUP> The current page is the list sweep setup page.
<TRACE SWEEP SETUP> The current page is the trace sweep setup page.
<TRACE SWEEP> The current page is the trace sweep display page.
<SYSTEM SETUP> The current page is the system setup page.
<FILE LIST> The current page is the file list page.
<TRANS TEST SET> The current page is the transformer test setup page.
<TRANS LIMIT SET> The current page is the transformer limit setup page.
<TRANS MEAS DISP> The current page is the transformer measurement
display page.
<TRANS JUDGE DISP> The current page is the transformer judge display
page.
The :LINE command is used to set the current measurement item which can be a
substring with up to 16 characters. The :LINE? query returns the current
measurement item. The character string of the measurement item can be used as
the file name when saving a file.
Command syntax: DISPlay:LINE<string>
Where,
<string> can be an ASCII character string (maximum number is 16).
For example: WrtCmd(DISP:LINE Resistor meas)
Query syntax: DISPlay:LINE?
Return format: <string><NL^END>
The :ResultFONt command is used to set the current font of the measurement
result. The :ResultFONt? Query returns the current font of the measurement
result.
Command syntax: DISPlay:RFONt <font>
<font> can be the following information:
LARGe: Use large character to display the measurement result, 12ms/meas.