TH2829X Series Operation Manual Chapter 10 HANDLER
178
The CORRection:LENGth command is used to set the correction cable length. The
CORRection:LENGth? query returns the current set cable length.
Command syntax: CORRection:LENGth<value>
Where,
<value> is 0, 1, 2 or 4 followed by M.
For example: WrtCmd (CORR:LENG 1M) Set the cable length as 1 meter.
Query syntax: CORRection:LENGth?
Return format: <NR1><NL^END>
The CORReciton:METHod command is used to set the correction mode. The
CORRection:METHod? query returns the current correction mode.
Command syntax: CORRection:METHod SINGle
MULTi
Where,
SINGle Set or return single channel mode.
MULTi Set or return multi channel mode.
For example: WrtCmd (CORR:METH MULT) Set the instrument as multi
channel mode.
Query syntax: CORRection:METHod?
Return format: SINGle <NL^END>
MULTi
The CORReciton:OPEN command is used to execute open correction for
48(TH2829C/CX) preset test points (TH2829AX has 41 preset test points,
TH2829A has 43 preset test points, TH2829B/BX has 45 preset test points).
Command syntax: CORRection:OPEN
For example: WrtCmd (CORR:OPEN)
The CORRection:OPEN:STATe command is used to set the open correction ON or
OFF. The CORReciton
OPEN:STATe? query returns the current open correction
status.
Command syntax:
ON
CORRction:OPEN:STATe OFF
1
0
Where,
1 (decimal 49) is equal to ON.
0 (decimal 48) is equal to OFF.
For example: WrtCmd (CORR:OPEN: STAT ON)
Query syntax: CORRection:OPEN:STATe?
Return format: <NR1><NL^END>