TH2829X Series Operation Manual Chapter 10 HANDLER
179
The CORRection:SHORt command is used to execute short correction for
48(TH2829C/CX) preset test points (TH2829AX has 41 preset test points,
TH2829A has 43 preset test points, TH2829B/BX has 45 preset test points).
Command syntax: CORRection:SHORt
For example: WrtCmd (CORR:SHOR)
The CORRection:SHORt:STATe command is used to set the short correction status.
The CORRection:SHORt:STATe? query returns the current short correction status.
Command syntax
ON
CORRection:SHORt:STATe OFF
1
0
Where,
1 (decimal 49) is equal to ON.
0 (decimal 48) is equal to OFF.
For example: WrtCmd (CORR:SHOR:STAT ON)
Query syntax: CORRection:SHORt:STATe?
Return format: <NR1><NL^END>
The CORRecition:LOAD:STATe command is used to set load correction. The
CORRection:LOAD:STATe? query returns the current load correction status.
Command syntax:
ON
CORRection:LOAD:STATe OFF
1
0
Where,
1 (decimal 49) is equal to ON.
0 (decimal 48) is equal to OFF.
For example: WrtCmd (CORR:LOAD:STAT ON)
Query syntax: CORRection:LOAD:STATe?
Return format: <NR1><NL^END>
The CORRection:LOAD:TYPE command is used to set the tested parameter type.
The CORRection:LOAD:TYPE query returns the current parameter type.
Details of function are as follows:
CPD Set the function as Cp-D LPRP Set the function as Lp-Rp
CPQ Set the function as Cp-Q LSD Set the function as Ls-D
CPG Set the function as Cp-G LSQ Set the function as Ls-Q
CPRP Set the function as Cp-Rp LSRS Set the function as Ls-Rs