TH2829X Series Operation Manual Chapter 10 HANDLER
ON
TRAN:TURN:STAT OFF
1
0
Where,
1 (decimal 49) is equal to ON.
0 (decimal 48) is equal to OFF.
For example: WrtCmd (“TRAN:TURN:STAT:ON”) All test parameter setting of
turn-ratio is valid.
Query syntax: TRAN:TURN:STAT?
Return format: <NR1><NL^END>
The TRAN:TURN:FREQ command is used to set the test frequency of turn-ratio.
The TRAN:TRUN:FREQ? query returns the current test frequency of turn-ratio.
Command syntax: TRAN:TURN:FREQ <value>
Where,
<value> can be NR1, NR2 or NR3 data format followed by HZ, KHZ, and MHZ.
For example: WrtCmd (“TRAN:TURN:FREQ 1KHZ”) Set the test frequency of
turn-ratio as 1KHZ.
Query syntax: TRAN:TURN:FREQ?
Return format: <NR3><NL^END>
The TRAN:TURN:LEVel command is used to set the test level of the turn-ratio.
The TRAN:TURN:LEVel query returns the current test level of the turn-ratio.
Command syntax: TRAN:TRUN:LEVel<value>
Where,
<value> can be NR1, NR2 or NR3 data format followed by mV or V.
NOTE: <value> ranges from 5mV to 10V. Beyond this range, error information
will be reported.
For example: WrtCmd (“TRAN:TURN:LEVel 1V”)
Query syntax: TRAN:TURN:LEVel?
Return format: <NR3><NL^END>
The TRAN:TURN: LImit command is used to set the primary nominal value,
secondary nominal value, high and low limits of secondary turn. The
TRAN:TURN:LImit query returns the current primary nominal value, secondary
nominal value, high and low limits of secondary turn.
Command syntax: TRAN:TURN:LIMit <Np value>,<Ns value>,<Ns low
limit>,<Ns high limit>
Where,