TH2829X Series Operation Manual Chapter 10 HANDLER
The TRAN:ACR:FREQ command is used to set the ACR test frequency. The
TRAN:ACR:FREQ? query returns the current ACR test frequency.
Command syntax: TRAN:ACR:FREQ<value>
Where,
<value> is NR1, NR2 or NR3 data format followed by HZ, KHZ, or MHZ.
For example: WrtCmd (“TRAN:ACR:FREQ 1KHZ”) Set the ACR test frequency as
1KHZ.
Query syntax: TRAN:ACR:FREQ?
Return format: <NR3><NL^END>
The TRAN:ACR:LEVel command is used to set the ACR test level. The
TRAN:ACR:LEVel query returns the current ACR level.
Command syntax: TRAN:ACR:LEVel<value>
Where,
<value> is NR1, NR2 or NR3 data format followed by mV or V.
NOTE: <value> ranges from 5mV to 2V. Beyond this range, error
information will be reported.
For example: WrtCmd (“TRAN:ACR:LEVEL 1V”) Set the ACR test level as 1V.
Query syntax: TRAN:ACR:LEVel?
Return format: <NR3><NL^END>
The TRAN:Cx:STAT command is used to set the Cx test parameter to be ON or OFF.
The TRAN:Cx:STAT query returns the current state of the Cx test parameter.
Command syntax:
ON
TRAN:Cx:STAT OFF
1
0
Where,
1 (decimal 49) is equal to ON.
0 (decimal 48) is equal to OFF.
For example: WrtCmd (“TRAN:Cx:STAT:ON”) Set the Cx test parameter to be
valid.
Query syntax: TRAN:Cx:STAT?
Return format: <NR1><NL^END>
The TRAN:Cx:FREQ command is used to set the Cx test frequency. The
TRAN:Cx:FREQ? query returns the current Cx test frequency.
Command syntax: TRAN:Cx:FREQ <value>
Where,