Program Mode 6-11
Connect
The DMF (Differential Mass Flow) test type includes a Connect step used to
pressurize the test port. Once the pressure is within the preset limits (indicating a
component has been connected), pressurizing stops before proceeding with the test
cycle.
The default regulator for the connect step is Regulator 1. This can be changed if an
additional regulator is fitted by clicking on Connect step, under the Program Mode,
Test Setup, Steps tab. Note if the step time elapses, the cycle aborts, indicating no
component has been connected and no further testing is required.