Program Mode 6-21
The TC Method selections are:
Disabled: Temperature compensation is disabled for the selected program
and associated parameters are hidden.
Enabled: The temperature compensation test steps are performed and the
compensation is applied.
Apply Only: The temperature compensation information from a previous
test is applied. This is generally used when programs are performed in a
sequence of tests and the first atmospheric reading is used by subsequent
tests in the sequence.
Steps Only: Performs the temperature compensation test steps but
compensation is not applied. This option is generally used during the initial
setup of temperature compensation to record values.
If tests are performed by individual programs then set the TC method to
Steps Only for all programs requiring Temperature compensation. If tests
are performed by a sequence of programs then set the TC Method to Steps
Only on the first program within the sequence.
(See page 8-16 for additional temperature compensation information.)
Challenge Setup Tab
• Use the Challenge Setup screen to set the parameters to use for the
challenge under leaking and non-leaking conditions.