<5. Functions>
103
IM 01E21A02-03EN
■Warning
Error XD_ERROR BLOCK_ERR
Process
value*
Totalizer Display
No Message STB DTB MTB LTB STB DTB MTB LTB
82
Pulse output 1
saturated
-- -- 17 -- -- -- Other --
Normal
calculation
Totalizer
processing
continue
Warning
85 Cable misconnect -- 17 -- -- -- Other -- --
Normal
calculation
Totalizer
processing
continue
Warning
86
Coil insulation
warning
-- 17 -- -- -- Other -- --
Normal
calculation
Totalizer
processing
continue
Warning
87
Adhesion over
level 3
17 17 -- -- Other Other -- --
Normal
calculation
Totalizer
processing
continue
Warning
88
Low conductivity
Warning
-- 17 -- -- -- Other -- --
Normal
calculation
Totalizer
processing
continue
Warning
89
Insulation
detection
-- 17 -- -- -- Other -- --
Normal
calculation
Totalizer
processing
continue
Warning
90
Flow noise over
level 3
-- 17 -- -- -- Other -- --
Normal
calculation
Totalizer
processing
continue
Warning
91
Flow noise over
level 4
17 17 -- -- Other Other -- --
Normal
calculation
Totalizer
processing
continue
Warning
92 Autozero warning 18 -- -- -- Other -- -- --
Normal
calculation
Totalizer
processing
continue
Warning
93
Verication
warning
-- -- -- -- -- -- -- --
Normal
calculation
Totalizer
processing
continue
Warning
94
Factory noise
warning
-- 17 -- -- -- Other -- --
Normal
calculation
Totalizer
processing
continue
Warning
95 Simulation active -- -- 17 -- -- -- Other --
Normal
calculation
Totalizer
processing
continue
Warning
98
Pulse output 1
xed
-- -- 17 -- -- -- Other --
Normal
calculation
Totalizer
processing
continue
Warning
101
Parameter restore
running
-- -- -- -- -- -- -- --
Normal
calculation
Totalizer
processing
continue
Warning
102
Display over
warning
-- -- -- 17 -- -- -- Other
Normal
calculation
Totalizer
processing
continue
Warning
103
microSD card size
warning
-- -- -- 17 -- -- -- Other
Normal
calculation
Totalizer
processing
continue
Warning
104
Parameter
backup
incomplete
-- -- 17 -- -- -- Other --
Normal
calculation
Totalizer
processing
continue
Warning
105
microSD card
mismatch
-- -- -- 20 -- -- -- Other
Normal
calculation
Totalizer
processing
continue
Warning
106
microSD
card removal
procedure error
-- -- -- 20 -- -- -- Other
Normal
calculation
Totalizer
processing
continue
Warning
131
Transmitter type
mismatch
19 -- -- -- Other -- -- --
Normal
calculation
Totalizer
processing
continue
Warning

Table of Contents

Other manuals for YOKOGAWA ADMAG TI Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA ADMAG TI Series and is the answer not in the manual?

YOKOGAWA ADMAG TI Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelADMAG TI Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals