<5. Functions>
52
IM 01E21A02-03EN
Display Devicesetup►Detailedsetup►Fieldbusinfo►AI1FB*►(seebelow)
F
OUNDATION
Fieldbus
DeviceConguration►AI1*►DeviceConguration►Conguration►BasicSetting►
(seebelow)
*For AXG, one from Al1 to Al4, and for AXW, one from AI1 to AI3.
Block Name
Relative
Index
Parameter
Description
Display F
OUNDATION Fieldbus
AI1
-AI4(AXG)
AI1
-AI3(AXW)
15 Channel CHANNEL
Selects CHANNEL number
used in the AI function block.
Refer to Table 5.1.2.1 for
selection.
ByselectingthechannelselectedinAI1,aprocessisdisplayedasPV(PrimaryVariable)in
Selected Flow for STB.
Display Devicesetup►Detailedsetup►Provar►(seebelow)
FOUNDATION
Fieldbus
DeviceConguration►STB►DeviceConguration►Conguration►Process
Variables►(seebelow)
Block
Name
Relative
Index
Parameter
Description
Display F
OUNDATION Fieldbus
STB 45 PVowselect Selected Flow
Displays the target process for
PV.
Settingexample:Whenthemagneticowmeterisusedbyoutputtingthevolumetricowrate
fromAI1andsettingthevolumetricowratespanat100m
3
/h, set the parameters as shown
below.
AI1: CHANNEL=“Volumeow”
AI1: TransducerScale.UnitsIndex=“m
3
/h”
AI1: TransducerScale.EUat100%=“100.000”
AI1: TransducerScale.EUat0%=“0.0”
STB:VolumeFlowSpan=“100.0”
Settingexample:Whenthemagneticowmeterisusedbyoutputtingthemassowratefrom
AI1andsettingthemassowratespanat10000kg/handthedensityat1000kg/m
3
, set the
parameters as shown below.
STB:Densityunit=“kg/m
3
”
STB:FixedDensity=“1000.0”
AI1: CHANNEL=“Massow”
AI1: TransducerScale.UnitsIndex=“kg/h”
AI1: TransducerScale.EUat100%=“10000.0”
AI1: TransducerScale.EUat0%=“0.0”
STB:MassFlowSpan=“10000.0”