2000X energy Power Supply Chapter 6: Operation
Instruction Manual Setting Primary Parameters
011-003-991 EN 6-17
Setting Weld Time
You can select the length of time (in seconds) that ultrasonic energy will be transmitted to
your parts.
Take this action… To get this result…
Press the Select key that corresponds to
Weld Time(s).
Use the numeric keypad to enter the Weld
Time (in seconds) that you want to use, or
press the same Select key repeatedly to
toggle between the minimum (0.010 sec-
ond) and maximum (30.000 seconds) val-
ues for Weld Time.
Press Enter on the numeric keypad to save
the new value, or press Go Back/Esc to
make no change to the value.
WELD SETUP Pg1of13
Weld Mode> Time
Weld Time(s)= 0.010
Hold Time(s)= 0.010
WELD SETUP Pg1of13
Weld Mode> Time
Weld Time(s)= 1.500
Hold Time(s)= 0.010
WELD SETUP Pg1of13
Weld Mode> Time
Weld Time(s)= 1.500
Hold Time(s) = 0.010