2000X energy Power Supply Chapter 7: Maintenance
Instruction Manual System Alarm Tables
011-003-991 EN 7-45
7.5.8 Note Alarms
In addition to the previously described alarms there are several Note alarms that the Controls can issue to
alert you that an alarm is imminent, or that the cycle ran with authorized modifications.
2000a
2000f
2000e
2000t
Seek Overload Seek Overload
The ultrasonic power supply
was overloaded during the
seek cycle at power up.
Check the stack.
Repair/Replace the
Power Supply Module.
For digital UPS only,
check that the stack is
properly attached and
the RF cable is hooked
up.
2000a
2000f
2000e
2000t
Test Overload Test Overload
The ultrasonic power supply
was overloaded during the test
cycle. Pressing Test will clear
the overload before the power
supply goes into test mode. A
new cycle can’t be run until
reset is pressed, even though
test can be pressed again.
Check the stack.
Repair/Replace the
Power Supply Module.
For digital UPS only,
check that the stack is
properly attached and
the RF cable is hooked
up.
Table 7.9 Overload Alarms
Model Display Message Printer Message Alarm Condition/Cause Corrective Action