ICS-3000 Ion Chromatography System
236 Doc. 065031-04 1/08
Chromeleon and Chromeleon Xpress compare the current date with the
column expiration date during injection or before beginning batch processing.
This error occurs if the expiration date has already passed or if batch
processing is scheduled to start less than 24 hours before the expiration date.
To troubleshoot:
Replace the column.
This error occurs if, during a firmware (Moduleware) download, the existing
firmware cannot be erased from memory.
To troubleshoot:
Flash memory cannot be serviced by the user. Contact Dionex for assistance.
This error occurs if the column ID chip card is installed incorrectly, or is
defective.
To troubleshoot:
When the chip card is properly installed, the Dionex logo on the card faces up
and the LED on the memory slot is green. If this error occurs when the chip
card is properly installed, it indicates that the card is defective and should be
replaced (for instructions, refer to Section 9.26
).
Either the current date is later than the expiration date for x
or the program or batch will end later than 24 hours before
the expiration time for x.
Error deleting flash memory.
Error during storing of the card data to the chip card.