ICS-3000 Ion Chromatography System
242 Doc. 065031-04 1/08
This error occurs if you select a TC temperature setting above 70 °C (158 °F).
To troubleshoot:
Refer to the column manual for the recommended operating conditions.
Column manuals are provided on the Dionex Reference Library CD-ROM
(P/N 053891).
If you select a value outside the allowable range for certain parameters,
Chromeleon ignores the setting, replaces it with the closest valid setting, and
this error occurs.
To troubleshoot:
No immediate action is required. In future, specify a valid setting for the
parameter.
Operating columns above 70 °C could affect column
performance. Verify recommended column operating
conditions before setting temperature above 70 °C.
Parameter x value exceeded its allowed range. Set to closest
valid value x.
-or-
Value for [name] parameter is not within valid range. Value
passed (x) will be substituted by y.
-or-
Value x replaced by valid table entry y.