TC Troubleshooting
Doc. 065031-04 1/08 243
This error occurs if you attempt to change one of the primary column
properties (batch or lot number, date of manufacture, etc.) used for column
identification.
To troubleshoot:
Once a primary column property has been assigned, it cannot be changed. As
an alternative, consider changing a secondary column property. For a list of
primary and secondary properties, refer to the Chromeleon or Chromeleon
Xpress Help.
This error occurs if you attempt to change column ID properties when at least
one column ID chip card is inserted incorrectly.
To troubleshoot:
1. In the Commands dialog box in Chromeleon or Chromeleon Xpress, click
the plus sign beside TC to display the items underneath.
2. One at a time, click the plus sign beside each column name and select the
CardState command. If the status is BusBlocked, remove the card from
the card slot and reinsert it.
When the card is properly installed, the Dionex logo on the card faces up
and the LED on the memory slot is green.
Primary property “x” is already assigned and cannot be
changed.
Properties cannot be changed. A chip card has not been
inserted properly. Check the CardState property for all cards
and reinsert those indicating “BusBlocked.”