ICS-3000 Ion Chromatography System
244 Doc. 065031-04 1/08
This error occurs if you attempt to change column ID properties when the TC
door is open (or is not fully closed).
To troubleshoot:
1. Verify that the door is fully closed.
2. If the door is not fully closed, check for—and remove—any obstruction.
3. If the door is fully closed but the error persists, contact Dionex for
assistance.
This error occurs if you attempt to change column ID properties for a column
for which no chip card is installed.
To troubleshoot:
Insert a column ID chip card into the appropriate slot (see Section 9.26
).
If a network failure prevents Chromeleon from creating or writing to a raw
data file, the corresponding error occurs.
To troubleshoot:
Wait until batch processing is complete; then, reboot the PC and restart the
Chromeleon server.
Properties cannot be changed if the compartment door is
open.
Properties cannot be changed—there is no card in the card
reader.
Raw data file x cannot be created.
-or-
Raw data file x cannot be created. Continuing with network
failure protection.
-or-
Raw data file x cannot be written.
-or-
Raw data file x cannot be written. Continuing with network
failure protection.