ICS-3000 Ion Chromatography System
246 Doc. 065031-04 1/08
This error occurs if you attempt to operate the TC when the door is open (or
not fully closed).
To troubleshoot:
1. Verify that the door is fully closed.
2. If the door is not fully closed, check for—and remove—any obstruction.
3. If the door is fully closed but the error persists, contact Dionex for
assistance.
This error occurs if the number of hours a TC part has been in use exceeds the
lifetime set by the OperationTime.Limit property.
To troubleshoot:
Replace the part identified in the message before resuming operation.
This error occurs if the column ID chip card data is inadvertently saved in an
invalid format.
To troubleshoot:
1. In the Commands dialog box in Chromeleon or Chromeleon Xpress, click
the plus sign beside TC to display the items underneath.
2. Click the plus sign beside the column name.
3. Select FormatChipCard and click Execute to reinitialize the chip card.
The compartment door is open; a batch or a program may
need to wait until the door has been closed. Please close the
door.
The counter [name] (value: x) has exceeded its limit (y).
Module should no longer be used. Replace the [name].
The data format stored on the chip card is unsupported
(format version x).