ICS-3000 Ion Chromatography System
248 Doc. 065031-04 1/08
Each time the TC power is turned on, the firmware (Moduleware) runs a
series of self-diagnostics in which the status of the main components is
checked. This error occurs if a component is not ready for operation.
To troubleshoot:
1. In the Commands dialog box in Chromeleon or Chromeleon Xpress, click
the plus sign beside TC to display the items underneath.
2. Check the status of each command listed in the error message. The table
below indicates the status required for operation.
3. When you identify the component that is not ready for operation, find and
eliminate the cause of the problem.
This error occurs when you modify a primary column property (batch or lot
number, date of manufacture, etc.). To prevent any unintended changes,
Chromeleon or Chromeleon Xpress requests confirmation before saving the
new property.
To troubleshoot:
If you are sure that the proposed change is correct, go ahead and save it.
The oven is not fully operable yet. Check the values of
“Standby,” “Ready,” “Door,” and “CardState.”
Command Status
CardState OK
Door Closed
Ready Ready
Standby NoStandby
The primary column ID property “x” will be changed. The
property can't be changed in the future.