TC Troubleshooting
Doc. 065031-04 1/08 249
This error occurs if the value entered for a parameter’s upper limit (for
example, TC_Temp.Signal.UpperLimit) is less than the lower limit for the
parameter.
To troubleshoot:
Increase the value of the upper limit setting.
This error occurs if the Warning value selected for a Predictive Performance
command (for example, LeftRotorSealSwitches) is lower than the Limit
value. Chromeleon or Chromeleon Xpress issues a warning message in the
Audit Trail when the Warning value is reached and an error message when
the Limit value is reached.
To troubleshoot:
Select a Warning value that is higher than the Limit value.
This error occurs if the Warning value selected for a parameter is higher than
the Limit value. Chromeleon or Chromeleon Xpress issues a warning
message in the Audit Trail when the Warning value is reached and an error
message when the Limit value is reached.
To troubleshoot:
Select a Warning value that is lower than the Limit value.
The upper limit must be higher than the lower limit.
The Warning threshold must be higher than the Limit
threshold.
The Warning threshold must be lower than the Limit
threshold.