ICS-3000 Ion Chromatography System
250 Doc. 065031-04 1/08
A Chromeleon or Chromeleon Xpress program can include no more than one
command to start or terminate data acquisition for a channel; if multiple
commands are present, the appropriate message (shown above) appears.
To troubleshoot:
Remove the duplicate commands from the program.
This error occurs if you attempt to change the setting for a read-only
parameter.
To troubleshoot:
Do not attempt to reset read-only parameters.
If batch processing is in progress when you attempt to change a property
(such as Temperature or TempCtrl) that affects the entire TC, this error
occurs.
To troubleshoot:
To avoid disrupting operation, cancel the command.
There was already an AcqOff command for this channel
before.
-or-
There was already an AcqOn command for this channel
before.
-or-
Two AcqOn commands for the same channel. The second
chromatogram will overwrite the first.
This function cannot be adjusted by the user.
This operation may affect the oven's operation on timebase
x, which is currently running a batch.