GE HEALTHCARE
DIRECTION 5394152, Revision 5
LOGIQ™ P6/P6 PRO SERVICE MANUAL
Page 5-36 Section 5-1 - Overview
5-8-2 System Checkout Procedure (cont’d)
All bold fields should be filled
7.) Device Name and CRM No are pre-populated and can not be changed.
8.) Select Continent and Country.
9.) Put City, State, and Institution.
10.)Select “Product” for the Enterprise.
11.)If Proxy needs to be configured, select “Enable” for the Proxy filed and input IP address and Port
number.
12.)Press “Submit Changes”. “Please Wait” message is displayed while the system is checking out.
Figure 5-37 Please Wait