GE HEALTHCARE
DIRECTION 5394152, Revision 5
LOGIQ™ P6/P6 PRO SERVICE MANUAL
Page 8-144 Section 8-4 - Software Installation
8-6-2-1 Perform Software Reloading in the Service Platform (cont’d)
3.) Press “Time Zone”
4.) Register the current Time Zone
5.) Press “Cancel”.
6.) Log onto the Service Platform.
7.) Go to “Utilities”->”Common Utilities”.
8.) Go to “Utility”->”System”->”About”
Verify that Software version and software build version are same with the ones before the Software
Reloading.
Figure 8-216 Time Zone
Current Time Zone
Table 8-1 Current Time Zone