EasyManua.ls Logo

GE Vivid S60

GE Vivid S60
709 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Measurements and Analysis
8-22 Vivid S70 / S60User Manual
BC092760-1EN
01
AFI Stage Menu
The workflow in AFI is controlled by the stage menu to the right
in the tool window, see Figure 8-12. It is possible to navigate
between the stages by clicking on the stage buttons. Stages that
are not accessible from the given tool state are greyed out.
Figure 8-12. AFI Stage menu
The stage menu buttons also contain information about
completeness state and a reset button.
By clicking the (reset) button, the stage is reset and all
user entered information in that stage is cleared. Any
automatic procedures are re-run.
If a stage is labeled , it means the stage is complete and
nothing more needs to be done in that tool stage.
Beneath the stage menu button, a helpstring for the current
active state is shown.
AFI on the APLAX view
1. Pay attention to the left/right orientation of the image by
comparing the LV wall names with a visual inspection of the
image. If the image orientation is wrong:
go back to the Select view stage.
press Left-Right Flip.
verify the view by annotating it as APLAX again.
NOTE: You may alternatively exit AFI, invert the image and start
AFI again.
2. An automatic ROI is generated when entering the Define
ROI stage, see Figure 8-10. The ROI may be edited by
clicking and dragging on the endocardial and epicardial
contours. See ‘ROI adjustment’ on page 8-25 for more
information.

Table of Contents

Other manuals for GE Vivid S60

Related product manuals