Customize Configuration
12-46 Vivid S70 / S60 – User Manual
BC092760-1EN
01
• Check Allow SR if required (see ‘DICOM SR’ on
page 12-48).
• Keep Reopen per image unchecked.
• Set Max Frame rate to 30.
• Keep Only Black and White unchecked.
• Set Compression to JPEG.
• Set Quality to 95.
• Check Allow multiframe.
• Raw Data Settings gives you the ability to select which
images to transfer in raw data format by mode:
Figure 12-31. Raw Data Settings window
• Systole only (for Stress). When this setting is active,
all stress images will be sent to the DICOM server
showing only systole. The setting is selectable when
transferring only multiframe (not raw data).
NOTE: Setting Compression to None may result in long
transfer time and cineloop with more than 500 frames
will be truncated. If Compression is set to None, set
Max frame rate to either 25 or 30 frames per second to
reduce the risk of truncating loops.
Adjusting the Search Criteria
When selecting a DICOM Worklist dataflow or Query/Retrieve,
search criteria can be set for the system to use when searching
the database.